Reduction Method of Patients' Radiation Dose Considering the Size of Field of View with a Digital Cine X-ray System Loading a Flat-panel Detector
Flat−panel Detector搭載型血管撮影装置における視野サイズを考慮した患者被ばく線量低減法
2011
Japanese Journal of Radiological Technology
Flat−panel Detector搭載型血管撮影装置における視野サイズを考慮した患者被ばく線量低減法
Angiographic equipment has tended toward using flat-panel detector (FPD). However, the radiation dose varies depending on the size of field of view (FOV) in the FPD. Although we were using 7 inch FOV in cardiac procedures, we examined a way to reduce the patients' radiation dose by using 8 inch FOV utilizing a digital zoom processing. We measured the absorbed dose rates at an interventional reference point (IRP) with 20 cm acrylic board to compare 8 inch FOV using a digital zoom processing with