De-airing method of an axial flow left ventricular assist device influences post-operative lactate dehydrogenase levels: a possible explanation for some episodes of pump thrombosis
2018
The VAD Journal: The Journal of Mechanical Assisted Circulation and Heart Failure
Pump thrombosis (PT) is a relatively uncommon but serious complication of a left ventricular assist device (LVAD). We believe that de-airing of the HeartMate II (HMII) with the pump turned on and a clamp across the outflow graft may lead to early thrombus formation due to heat generation on the bearings. Methods Patients who underwent HMII implantation from November 2012 to February 2016 were retrospectively reviewed. Patients were separated into two groups depending on the timing of removing
doi:10.14434/vad.v4i0.28045