Radiobiological evaluation considering setup error on single‐isocenter irradiation in stereotactic radiosurgery
2021
Journal of Applied Clinical Medical Physics
We calculated the dosimetric indices and estimated the tumor control probability (TCP) considering six degree-of-freedom (6DoF) patient setup errors in stereotactic radiosurgery (SRS) using a single-isocenter technique. We used simulated spherical gross tumor volumes (GTVs) with diameters of 1.0 cm (GTV 1), 2.0 cm (GTV 2), and 3.0 cm (GTV 3), and the distance (d) between the target center and isocenter was set to 0, 5, and 10 cm. We created the dose distribution by convolving the blur component
doi:10.1002/acm2.13322
pmid:34151498
fatcat:k72sxkunj5gmnfihgw4iy3hz5e