Separation of Small DNAs by Gel Electrophoresis in a Fused Silica Capillary Coated with a Negatively Charged Copolymer

2017 Separations  
Active development of compact analytical instruments suitable for point-of-care testing (POCT) requires optimization of existing methods. To aid the development of capillary gel electrophoresis instruments for POCT, we attempted to separate polymerase chain reaction products (small DNAs) using a short, fused silica capillary coated with an acrylamide (AM)/acrylic acid (AA) copolymer (poly(AM-co-AA)). To realize the high capability of this capillary to separate small DNAs, the magnitude of
more » ... magnitude of electroosmotic flow (EOF) was controlled by varying the content of negatively charged AA in the copolymer, which significantly affected the separation ability. At an AA content ≥3.75 mol %, sample DNAs could not be injected into the copolymer-coated capillary owing to strong EOF, whereas a 100 bp DNA ladder sample was successfully separated at an AA content of ≤3.5 mol %, showing that even slight AA content variations impact DNA flow. EOF values measured using a neutral coumarin 334 solution suddenly decreased at an AA content of 3.5 mol % relative to those at an AA content of ≥3.75 mol %. Theoretical plate values revealed that an AA content of 2.75 mol % was optimal for separating ladder DNAs with sizes ≥600 bp. Hence, EOF control achieved by varying the amount of negatively charged AA in the poly(AM-co-AA) coating can promote further development of short capillaries for POCT applications.