A Compact Slot-loaded Antipodal Vivaldi Antenna for a Microwave Imaging System to Monitor Liver Microwave Thermal Ablation
2022
IEEE Open Journal of Antennas and Propagation
This study presents the design and the experimental validation of a slot-loaded antipodal Vivaldi antenna. The intended use is in an array configuration for monitoring liver microwave thermal ablation by way of microwave imaging (MWI). To optimize electromagnetic power transfer to the human abdomen, the antenna was designed to operate in a coupling medium. The final design has overall dimensions of 40mm×65mm, and the working bandwidth goes from 600 MHz up to 3 GHz, with the possibility to