Radiometer standard for absolute responsivity calibrations from 950 nm to 1650 nm with 0.05% (k= 2) uncertainty
2009
Metrologia
A near-IR radiometer standard with similar performance to silicon trap detectors has been developed to calibrate detectors and radiometers for absolute spectral power, irradiance and radiance responsivities between 950 nm and 1650 nm. The new radiometer standard is utilized at the Spectral Irradiance and Radiance Responsivity Calibrations using Uniform Sources (SIRCUS) which is the reference calibration facility of NIST for absolute responsivity. The radiometer standard is a sphere detector