Calibrations of Hydraulic Pressure Transducers with Stepwise and 0-A-0-type Procedures
2012
Transactions of the Society of Instrument and Control Engineers
This study focuses on the effects of the calibration protocol on the results for hydraulic pressure transducers. Using a fully-automated calibration system that uses a pressure balance as the standard device, quartz Bourdon-type hydraulic pressure transducers were calibrated at pressures from 10 MPa to 100 MPa. Two kinds of calibration procedures, stepwise and 0-A-0-type procedures, were applied to pressure increasing and decreasing cycles. For the stepwise procedure, the interval between the