A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2020; you can also visit the original URL.
The file type is application/pdf
.
Comparative Analysis of G.M Tube Characteristics Using Ba-133 Gamma Source and Sr-90 Beta Test Source
2017
International Journal of Trend in Scientific Research and Development
The study is a recap of experimental physics which compares the G.M tube characteristics of (Ba-133) and (Sr-90) source respectively. SPECTECH ST360 Counter was used for this experiment. The test sources where placed 2cm and 4cm respectively in the counter window and voltage applied through starting at a potential of 780v. The counter was allowed to read for 60sec as the voltages was increased in step of 20v. Results obtained using Microsoft mathematics and further evaluations showed an
doi:10.31142/ijtsrd92
fatcat:odebndk5nrawfjhkpghajt24ie