Analysis and demonstration of atmospheric methane monitoring by mid-infrared open-path chirped laser dispersion spectroscopy

Nart S. Daghestani, Richard Brownsword, Damien Weidmann
2014 Optics Express  
Atmospheric methane concentration levels were detected using a custom built laser dispersion spectrometer in a long open-path beam configuration. The instrument is driven by a chirped distributed feedback mid-infrared quantum cascade laser centered at ~1283.46 cm −1 and covers intense rotational-vibrational transitions from the fundamental ν 4 band of methane. A full forward model simulating molecular absorption and dispersion profiles, as well as instrumental noise, is demonstrated. The
more » ... ent's analytical model is validated and used for quantitative instrumental optimization. The temporal evolution of atmospheric methane mixing ratios is retrieved using a fitting algorithm based on the model. Full error propagation analysis on precision gives a normalized sensitivity of ~3 ppm.m.Hz -0.5 for atmospheric methane.
doi:10.1364/oe.22.0a1731 pmid:25607487 fatcat:h443x4kvbzbtjlhry2voxj5yle