Optimization of Mass Resolution Parameters Combined with Ion Cooler Performance
2022
High mass resolution spectrometers (HRMS) for separation of exotic ion species in nuclear physics experiment request a low emittance and small energy spread (with D E the peak-to-peak value, and sE the rms value) of the input beam, so that ion cooler devices, as a Radio Frequency Quadrupole Coolers (RFQC), are typically envisioned. The SPES (Selective Production of Exotic Species) project at LNL requests M/(D M) about 20000, rms normalized emittance in the order of 2 nm, and for 160 keV ions,
doi:10.18429/jacow-ipac2022-thpotk003