A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2016; you can also visit the original URL.
The file type is application/pdf
.
Methodology of QT-Interval Measurement in the Modular ECG Analysis System (MEANS)
2009
Annals of Noninvasive Electrocardiology
Background. QT prolongation as can be induced by drugs, signals the risk of lifethreatening arrhythmias. The methodology of QT measurement in the Modular ECG Analysis System MEANS is described. Methods. In the simultaneously recorded leads of the standard 12-lead ECG, the QRS complexes are detected by a spatial velocity function. They are typed as dominant or non-dominant, and a representative complex per lead is obtained by averaging over the dominant complexes. QRS onset and T end are
doi:10.1111/j.1542-474x.2008.00261.x
pmid:19143743
fatcat:i3mvhag7vzfgho5kwkklnvcyuy