The effect of anatomical variables and use of the Lifts system on hearing outcomes after implantation of an active transcutaneous bone conduction device in bilateral congenital conductive hearing loss
2020
Journal of Otolaryngology - Head and Neck Surgery
Malformations of the temporal bone present different challenges to the implantation of a transcutaneous active bone conduction device, such as Bonebridge (Med-el, Innsbruck, Austria). This study aims to describe the benefits of high-resolution computed tomography (HRCT) in preoperative assessment and to analyze whether characteristics of the mastoid process, intraoperative compression of the dura or sigmoid sinus, and the use of the Lifts system, lead to differences in audiological performance