Oxygen-Sensing Paint-On Bandage

Pieter G. L. Koolen, Zongxi Li, Emmanuel Roussakis, Marek A. Paul, Ahmed M. S. Ibrahim, Robina Matyal, Thomas Huang, Conor L. Evans, Samuel J. Lin
2017 Plastic and Reconstructive Surgery  
Knowledge of tissue oxygenation status is fundamental in the prevention of postoperative flap failure. Recently, the authors introduced a novel oxygen-sensing paint-on bandage that incorporated an oxygen-sensing porphyrin with a commercially available liquid bandage matrix. In this study, the authors extend validation of their oxygen-sensing bandage by comparing it to the use of near-infrared tissue oximetry in addition to Clark electrode measurements. The oxygen-sensing paint-on bandage was
more » ... lied to the left hind limb in a rodent model. Simultaneously, a near-infrared imaging device and Clark electrode were attached to the right and left hind limbs, respectively. Tissue oxygenation was measured under normal, ischemic (aortic ligation), and reperfused conditions. On average, the oxygen-sensing paint-on bandage measured a decrease in transdermal oxygenation from 85.2 mmHg to 64.1 mmHg upon aortic ligation. The oxygen-sensing dye restored at 81.2 mmHg after unclamping. Responses in both control groups demonstrated a similar trend. Physiologic changes from normal to ischemic and reperfused conditions were statistically significantly different in all three techniques (p < 0.001). The authors' newly developed oxygen-sensing paint-on bandage exhibits a comparable trend in oxygenation recordings in a rat model similar to conventional oxygenation assessment techniques. This technique could potentially prove to be a valuable tool in the routine clinical management of flaps following free tissue transfer. Incorporating oxygen-sensing capabilities into a simple wound dressing material has the added benefit of providing both wound protection and constant wound oxygenation assessment.
doi:10.1097/prs.0000000000003421 pmid:28654595 fatcat:mwvh6vilfbb7xjaxu3eip47ohe