Solving Inaccuracies in Anatomical Models for Electrocardiographic Inverse Problem Resolution by Maximizing Reconstruction Quality

Miguel Rodrigo, Andreu M. Climent, Alejandro Liberos, Ismael Hernandez-Romero, Angel Arenal, Javier Bermejo, Francisco Fernandez-Aviles, Felipe Atienza, Maria S. Guillem
2018 IEEE Transactions on Medical Imaging  
Electrocardiographic Imaging (ECGI) has become an increasingly used technique for non-invasive diagnosis of cardiac arrhythmias, although the need for medical imaging technology to determine the anatomy hinders its introduction in the clinical practice. This work explores the ability of a new metric based on the inverse reconstruction quality for the location and orientation of the atrial surface inside the torso. Body surface electrical signals from 31 realistic mathematical models and four AF
more » ... patients were used to estimate the optimal position of the atria inside the torso. The curvature of the Lcurve from the Tikhonov method, which was found to be related to the inverse reconstruction quality, was measured after application of deviations in atrial position and orientation. Independent deviations in the atrial position were solved by finding the maximal L-curve curvature with an error of 1.7±2.4 mm in mathematical models and 9.1±11.5 mm in patients. For the case of independent angular deviations, the error in location by using the L-curve was 5.8±7.1º in mathematical models and 12.4º±13.2º in patients. The ability of the L-curve curvature was tested also under superimposed uncertainties in the 3 axis of translation and in the 3 axis of rotation and the error in location was of 2.3±3.2 mm and 6.4º±7.1º in mathematical models, and 7.9±10.7 mm and 10.0º±12.8º in patients. The curvature of Lcurve is a useful marker for the atrial position and would allow emending the inaccuracies in its location.
doi:10.1109/tmi.2017.2707413 pmid:28541896 fatcat:hslbjxjwg5abdiryj2ytcqoi3m