Spread of the Optical Power Emission of Several Units of the Same Laser Therapy Devices Used in Sports Medicine, Which Cannot Be Assessed by the Users, Shown by Means of High-Fidelity Laser Measurement Technology
[post]
2023
unpublished
Laser therapy devices (LTDs) operating with near-infrared laser light are increasingly being used in sports medicine. For several reasons the users cannot evaluate whether or not such devices emit laser beams according to the specifications provided by the manufacturer and the settings of the device. In this study the laser beams from two different LTDs that can be used in sports medicine were thoroughly characterized by measuring the emitted power, pulse shapes and lengths, and spatial
doi:10.20944/preprints202301.0251.v1
fatcat:qs4n5mkw3jdxdhjvry43zp56iy