ANALISIS KUALITAS DATA KELUARAN HARIAN ELEKTRON AKSELERATOR LINIER
2020
Komunikasi fisika Indonesia
Linac radiotherapy measurentoften experience instability. One of the irradiation errors with Linac can occur because the radiation beam that comes out is not expected properly. Determination of the correction factor and linearity is important to analyze the charge output of the electron energy emitted by Linac's modality to see the stability of the emitted charge. This study uses Electron Linear Accelerator Electron beam daily output data with a 10 cm × 10 cm applicator, 100 cm SSD, 0.125 cc