Evaluation of Various Scintillator Materials in Radiation Detector Design for Positron Emission Tomography (PET)
2020
Crystals
The performance of radiation detectors used in positron-emission tomography (PET) is determined by the intrinsic properties of the scintillators, the geometry and surface treatment of the scintillator crystals and the electrical and optical characteristics of the photosensors. Experimental studies were performed to assess the timing resolution and energy resolution of detectors constructed with samples of different scintillator materials (LaBr3, CeBr3, LFS, LSO, LYSO: Ce, Ca and GAGG) that were
doi:10.3390/cryst10100869