Adapting Raman Spectra from Laboratory Spectrometers to Portable Detection Libraries

James C. Weatherall, Jeffrey Barber, Carolyn S. Brauer, Timothy J. Johnson, Yin-Fong Su, Christopher D. Ball, Barry T. Smith, Rick Cox, Robert Steinke, Patricia McDaniel, Louis Wasserzug
2013 Applied Spectroscopy  
Raman spectral data collected with high-resolution laboratory spectrometers are processed into a format suitable for importing as a user library on a 1064 nm DeltaNu first generation, field-deployable spectrometer prototype. The two laboratory systems used are a 1064 nm Bruker Fourier transform (FT)-Raman spectrometer and a 785 nm Kaiser dispersive spectrometer. The steps taken to adapt for device-dependent spectral resolution, wavenumber shifts between instruments, and relative intensity
more » ... se are described. Effects due to the differing excitation laser wavelengths were found to be minimal, indicating-at least for the near-infrared (NIR)-that data can be ported between different systems, so long as certain measures are taken with regard to the reference and field spectra.