Motion and muscle artifact removal validation using an electrical head phantom, robotic motion platform, and dual layer mobile EEG

Natalie Richer, Ryan J. Downey, W. David Hairston, Daniel P. Ferris, Andrew D. Nordin
2020 IEEE transactions on neural systems and rehabilitation engineering  
Motion and muscle artifacts can undermine signal quality in electroencephalography (EEG) recordings during locomotion. We evaluated approaches for recovering ground-truth artificial brain signals from noisy EEG recordings. We built an electrical head phantom that broadcast four brain and four muscle sources. Head movements were generated by a robotic motion platform. We recorded 128-channel dual layer EEG and 8-channel neck electromyography (EMG) from the head phantom during motion. We
more » ... ground-truth electrocortical source signal recovery from artifact contaminated data using Independent Component Analysis (ICA) to determine: (1) the number of isolated noise sensor recordings needed to capture and remove motion artifacts, (2) the ability of Artifact Subspace Reconstruction to remove motion and muscle artifacts at contrasting artifact detection thresholds, (3) the number of neck EMG sensor recordings needed to capture and remove muscle artifacts, and (4) the ability of Canonical Correlation Analysis to remove muscle artifacts. We also evaluated source signal recovery by combining the best practices identified in aims 1-4. By including isolated noise and EMG recordings in the ICA decomposition, we more effectively recovered ground-truth artificial brain signals. A reduced subset of 32-noise and 6-EMG channels showed equivalent performance compared to including the complete arrays. Artifact Subspace Reconstruction improved source separation, but this was contingent on muscle activity amplitude. Canonical Correlation Analysis also improved source separation. Merging noise and EMG recordings into the ICA decomposition, with Artifact Subspace Reconstruction and Canonical Correlation Analysis preprocessing, improved source signal recovery. This study expands on previous head phantom experiments by including neck muscle source activity and evaluating artificial electrocortical spectral power fluctuations synchronized with gait events.
doi:10.1109/tnsre.2020.3000971 pmid:32746290 fatcat:dlgs6ansabgojjz55mb5e3a4vq