Interactions Between Electronic Article Surveillance Systems and Implantable Cardioverter-Defibrillators
1999
Circulation
Background-In patients with implantable cardioverter-defibrillators (ICDs). inappropriate shocks have been reported with exposure to electronic article surveillance systems. The risk to patients with ICDs of walking through or lingering near surveillance systems requires further investigation. Methods and Results-We evaluated the response in ICD function in 170 subjects during a 10-to 15-second midgate walk-through of and during extreme (2 minutes within 6 in of the gate) exposure to 3 common
doi:10.1161/01.cir.100.4.387
pmid:10421599