A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2021; you can also visit the original URL.
The file type is application/pdf
.
The Commissioning and Validation of EclipseTM Treatment Planning System on a Varian VitalBeamTM Medical Linear Accelerator for Photon and Electron Beams
2021
Frontiers in biomedical technologies
Purpose: Commissioning of a linear accelerator is a process of acquiring a set of data used for patient treatment. This article presents the beam data measurement results from the commissioning of a VitalBeamTM linac. Materials and Methods: Dosimetric properties for 6,10, and 15 MV photon beams and 6, 9, 12, and 16 MeV electron beams have been performed. Parameters, including Percentage Depth Dose (PDD), depth dose profile, symmetry, flatness, quality index, output factors, and the vital data
doi:10.18502/fbt.v8i2.6514
fatcat:qfcu7kdgmfatrnpzxiji4vhqme