Calibrating module for local hyperthermia complex «Phoenix-2»
Калибровочный модуль для комплекса локальной гипертермии «Феникс-2»
2016
Proceedings of Tomsk State University of Control Systems and Radioelectronics
Калибровочный модуль для комплекса локальной гипертермии «Феникс-2»
The objective of calibrating module is to heat needle heater and maintain its temperature on certain level with maximum deviation ±0.1 °C. The purpose is achieved using precision temperature sensor and proportional-integral-derivative controller in automatic control system. According the test results, the static error does not exceed 0.04 °C
doi:10.21293/1818-0442-2016-19-4-125-128
fatcat:4b6nett44re4jgtyfsr4z6rdgi