A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2018; you can also visit the original URL.
The file type is application/pdf
.
Fast Electron Thermometry for Ultrasensitive Calorimetric Detection
2015
Physical Review Applied
We demonstrate radiofrequency thermometry on a micrometer-sized metallic island below 100 mK. Our device is based on a normal metal-insulator-superconductor tunnel junction coupled to a resonator with transmission readout. In the first generation of the device, we achieve 90 μK/Hz^1/2 noise-equivalent temperature with 10 MHz bandwidth. We measure the thermal relaxation time of the electron gas in the island, which we find to be of the order of 100 μs. Such a calorimetric detector, upon
doi:10.1103/physrevapplied.3.014007
fatcat:r2eizjkcebbwbawtx35a2uxyxq