New ultraviolet radiometry beamline at the Synchrotron Ultraviolet Radiation Facility at NIST

P-S Shaw, K R Lykke, R Gupta, T R O'Brian, U Arp, H H White, T B Lucatorto, J L Dehmer, A C Parr
1998 Metrologia  
We have constructed a new ultrav iolet (U V) radiom etry facility at the Synchrotron Ultraviolet Radiation Facility (SURF II) at the National Institute of Standards and Technology (N IST). The facility combines a high-throu ghput norm al-in cidence monochrom ator with an absolute cryogenic radiom eter (A CR) optim ized for UV m easurem ents to provide absolute detector-based radiom etric calibratio ns in the spectral range 125 nm to approxim ately 320 nm. The system can be con® gured for
more » ... radiom etric calibration of photodetectors with completely automated m apping of detecto r spatial and angular responsivity . The facility can easily be adapted for other spectro radiom etric measurem ents includ ing tran smittan ce and re¯ectance and the spectral range can be extended to about 50 nm.
doi:10.1088/0026-1394/35/4/15 fatcat:wr5bnts4xvamfln5mdljhdgt6m