DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF OFLOXACIN AND ORNIDAZOLE IN PRESENCE OF ORNIDAZOLE IMPURITY IN COMBINED PHARMACEUTICAL DOSAGE FORM

Farrel Lisa Gauncar
2017 World Journal of Pharmaceutical Research  
A simple, rapid, accurate and precise RP-HPLC method has been developed for simultaneous estimation of Ofloxacin and Ornidazole and Ornidazole Impurity A in combined dosage form. The column used was Phenomenex C18 (250 mm×4.6 mm, 5µm) with a flow rate of 1.0 ml/min using UV detection at 318 nm. The chromatographic separation was carried out using mobile phase containing ACN: Phosphate buffer at pH 3.0 adjusted with OPA in the ratio of 40:60 v/v. The described method was validated and linearity
more » ... as observed over a concentration range of 2-20 µg/ml for Ofloxacin, 5-50 µg/ml for Ornidazole and 0.5-2 µg/ml for Impurity A with correlation coefficient of (R 2 ˃0.999). All components were well resolved within 5 minutes. The proposed method can be applied effectively during routine purity testing and successfully for determination of Ofloxacin, Ornidazole and Ornidazole Impurity A in bulk and pharmaceutical dosage form.
doi:10.20959/wjpr20179-9213 fatcat:ng6h52cm5bebffajdp5kakrivy