Improved assay protocol for measurement of ammonia on the Roche Cobas 8000 automated platform

J. Kaplon, J.J. de Groot, J.P. van Straalen, M. Heckman, J.C. Fischer
2019 Practical Laboratory Medicine  
Ammonia is a metabolite of protein catabolism that, when elevated, may be toxic for tissues, especially for the central nervous system. Elevated ammonia in blood is an indicator and a prognostic factor for hepatic and kidney disease or inherited metabolic disorders in nitrogen metabolism. The accuracy of ammonia determination is influenced by sampling condition, handling, storage and assay itself. Our and other laboratories have been experiencing high frequencies sample error flags while
more » ... ng ammonia with glutamate dehydrogenase method on Roche Cobas 8000 platform. To reduce the number of error flags we adapted Roche NH3L protocol by incorporation of an additional onboard routine step for sample pre-dilution. The AMC NH3L is an adaptation of Roche protocol that uses four fold pre-dilution of the sample in the rerun prior to the analysis. It was assessed for 1.occurrence of absorbance error flags, 2.precision, 3.correlation with Roche method and 4.interference by hemolysis, icterus and lipemia. The AMC NH3L adaptation demonstrates acceptable within-run and total precision. Comparison studies show no differences between the Roche rerun application and AMC NH3L adaptation. The AMC NH3L adaptation solves 78% of absorbance errors and for samples with high ammonia concentration is less affected by interferences from icterus and hemolysis than the Roche rerun application. The AMC NH3L adaptation is less prone to instrument error flags and for samples with high ammonia concentration, is more robust to endogenous interferences. The AMC NH3L adaptation is viable alternative to the Roche protocol for the ammonia measurement.
doi:10.1016/j.plabm.2018.e00115 pmid:30581950 pmcid:PMC6295606 fatcat:gvt43junpffzbibkelx7saps2i