A Comparative Study on the Lens Dose According to the Change of Shielding Material Used in Brain Computed Tomography
Brain CT에서 차폐 재료 변화에 따른 수정체 선량 비교 연구

Incheol Hwang, Woonjae Shin, Eunbo Gang
2015 Journal of the Korean Society of Radiology  
The cases of radiographic inspection for medical diagnosis in Korea have been continuously increasing year after year, which pays particularly more attention to CT which occupies over the half of medical radiation exposure. To find an effective alternative for reducing radiation exposure, the researchers conducted comparative experiments using some shields made of bismuth, aluminum 6mm, and silicone 22mm. These shielding materials have been used to reduce the entrance surface dose (ESD) on
more » ... s, maintaining the CT number, noise, and uniformity in brain CT scanning which forms the largest part in CT scanning these days. These experiments showed that the doses in the spiral scan parallel to IOML and the conventional scan in Bismuth were 26.41% and 17.52%, respectively; in Aluminum 18.24% and 9.39%; in Silicone 19.47% and 14.39% lower than compared with those in the cases without any shields. In the items of the CT number, noise, and uniformity, the bismuth shield satisfied exceedingly the standards of the phantom image test while aluminum and silicone were within. To keep the graphic quality and get good shielding effect, we recommend the silicone shield which can be manufactured and purchased with ease.
doi:10.7742/jksr.2015.9.1.31 fatcat:cy3a4aw5lzd2pppgmjk6xazqta