The "push-pull" dosimeter: When two pigments are better than one

Stephen G Bosi, Clive Baldock, Robert Smee
2010 Journal of Physics, Conference Series  
A new kind of gel dosimeter (the "push-pull" dosimeter) is proposed which would contain two spectrally complementary pigments, one which darkens with increasing dose and another which bleaches. The bleaching pigment would be optimised for high sensitivity and the darkening pigment for low sensitivity. By employing dual pigments optimised independently, the usual requisite compromises between sensitivity at low dose and accuracy at high dose would be relaxed. Such a gel, after exposure would be
more » ... ead using two successive optical CT scans employing two different scan wavelengths. The use of dual pigments could also reduce the occurrence of regions of high optical attenuation which generate artefacts in optical CT. This paper also presents results of simulations of the behaviour of such a gel when scanned using optical cone beam CT.
doi:10.1088/1742-6596/250/1/012010 fatcat:sedrfvcrcjer5pv7x7gmwkk4bu