A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2022; you can also visit the original URL.
The file type is application/pdf
.
Energy budget of an implantable glucose measurement system
2012
Biomedical Engineering
The theoretical evaluation of an implantable system measuring glucose for diabetes diagnostics and therapy has recently been presented. This work reports on the realisation of the proposed system on a printed circuit board and measurements of its energy budget. It is shown that the system can be run with small 3.2 V batteries for more than six months with an average energy consumption of 0.12 mAh.
doi:10.1515/bmt-2012-4445