完全体内埋め込み型生体信号モニタリング装置のための経皮デジタル通信と経皮電力供給システム
Transcutaneous Digital Communication and Energy Transmission for Total Implantable Biosignal Monitoring Device

Yusuke SAKAUE, Syunsuke ODA, Masahiko FUJITA, Yoshikazu AOKI, Kengo YAMANAKA, Eiichi HIRAI, Naruhiro SHIOZAWA, Masaaki MAKIKAWA
2009 Transactions of Japanese Society for Medical and Biological Engineering  
The final goal of this study is to develop a total implantable biosignal monitoring device which monitors biosignals inside body for long time. For such device transcutaneous digital communication and energy transmission is indispensable. This paper describes a noncontact transcutaneous digital communication using Bluetooth and long distance energy transmission system using core coil. We have developed an implantable biosignal monitoring device equipped with an acceleration sensor, an ECG
more » ... ier, a temperature sensor, a flash memory, a Bluetooth module, a transcutaneous energy receiving coil and a lithium ion battery. We performed an implant test using a cow and transcutaneous digital communication at a distance of 2, 700 mm and energy transmission at a distance of 150 mm through the skin were confirmed.
doi:10.11239/jsmbe.47.222 fatcat:7axxwnvptfg6bbzgpeoxcpj6wa