Staff dose calculation during a cobalt-60 source stuck

Eman Massoud, Hanan Diab
2006 Nuclear Technology and Radiation Protection  
One of the most pos tu lated ac ci dents in 60 Co ra dio ther apy units is the source get ting stuck, where one or more of the staff should en ter the treat ment room to deal with the prob lem. For such an ac ci dent, an emer gency plan is im por tant. A three-di men sional model of a 60 Co ther apy room has been done us ing the Monte Carlo code MCNP4B. The ra di a tion safety mea sures taken and the draw ings of the de vice are given to gether with sug ges tions for fu ture use of the source
more » ... use of the source for ir ra di a tion pur poses. More over, the cal cu lated re sults were com pared with those of an ex per i men tal study deal ing with this prob lem and were found to be in very good agree ment. Key words: source stuck, ra di a tion ex po sure in ra dio ther apy, dose cal cu la tion 48 Nu clear Tech nol ogy & Ra di a tion Pro tec tion -1/2006 Tech ni cal pa per UDC: 615.849.5:519.245