Performance Characteristics of the Abbott Architect SARS-CoV-2 IgG Assay and Seroprevalence Testing in Idaho [article]

Andrew Bryan, Gregory Pepper, Mark H. Wener, Susan L Fink, Chihiro Morishima, Anu Chaudhary, Keith Jerome, Patrick C Mathias, Alex Greninger
2020 medRxiv   pre-print
Coronavirus disease-19 (COVID19), the novel respiratory illness caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is associated with severe morbidity and mortality. The rollout of diagnostic testing in the United States was slow, leading to numerous cases that were not tested for SARS-CoV-2 in February and March 2020, necessitating the use of serological testing to determine past infections. Here, we evaluated the Abbott SARS-CoV-2 IgG test for detection of anti-SARS-CoV-2
more » ... IgG antibodies by testing 3 distinct patient populations. We tested 1,020 serum specimens collected prior to SARS-CoV-2 circulation in the United States and found one false positive, indicating a specificity of 99.90%. We tested 125 patients who tested RT-PCR positive for SARS-CoV-2 for which 689 excess serum specimens were available and found sensitivity reached 100% at day 17 after symptom onset and day 13 after PCR positivity. Alternative index value thresholds for positivity resulted in 100% sensitivity and 100% specificity. We then tested 4,856 individuals from Boise, Idaho collected over one week in April 2020 as part of the Crush the Curve initiative and detected 87 positives for a positivity rate of 1.79%. These data demonstrate excellent analytical performance of the Abbott SARS-CoV-2 IgG test as well as the limited circulation of the virus on the West Coast. We expect the availability of high-quality serological testing will be a key tool in the fight against SARS-CoV-2.
doi:10.1101/2020.04.27.20082362 fatcat:cclcbrg4pbgy5b3hzfn2co5o4a