MEASUREMENT OF VELOCITY DISTRIBUTION FOR FLOW THROUGH A BILEAFLET VALVE USING COLOR-CODED DIGITAL PARTICLE TRACKING VELOCIMETRY
2005
Journal of Marine Science and Technology
An experimental method, the so-called color-coded Digital Particle Tracking Velocimetry system was set up in this study to investigate viscous flow through an artificial bileaflet valve with three different opening angles, 5°, 30°, and 45°, respectively. An oscilloscope that generated stable wave speed was taken to calibrate measuring result of the equipment. Besides, a fully developed pipe flow case was used to check the reliability of experimental system, and a statistical t-test was
doi:10.51400/2709-6998.2098