A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2018; you can also visit the original URL.
The file type is application/pdf
.
Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform
2002
Europace
Conclusions The time constant of the membrane depends on the field, action potential phase and the shock polarity, but exceeds 1 msec. Therefore, we suggest using a slower shock leading edge, since the membrane cannot follow potentially damaging faster waveforms.
doi:10.1053/eupc.2001.0199
pmid:11846315