Safety and Efficacy of Magnet Use to Temporarily Inhibit Inappropriate Subcutaneous Implantable Cardioverter Defibrillator Therapy in Emergency Situations: A Case Report

Maurizio Santomauro, Carla Riganti, Mario Alberto Santomauro, Aniello Viggiano, Gaetano Castellano, Gianluigi Iovino, Antonio Rapacciuolo, Francesco Fiore, Francesco Cacciatore, Giovanni Esposito
2022 Journal of Cardiovascular Emergencies  
Introduction: The subcutaneous implantable cardioverter defibrillator (S-ICD) represents a major advancement in ICD technology. Inappropriate shocks (IAS) occur in more than 3.1% of the population with S-ICD each year and are usually followed by admission to the emergency department (ED). In this setting, the disabling of IAS is mandatory during a pseudo-electrical storm (ES). This report describes the strategies that can be followed in order to temporarily inhibit IAS in critical care settings
more » ... with the use of magnets. Case presentation: An S-ICD was implanted more than 6 weeks prior to presentation in a 68-year-old man with hypertrophic cardiomyopathy. In the ED, the patient experienced 3 IAS in the last hour. A Medtronic magnet was applied to stop IAS, as the specific programmer was not available. The maneuver interrupted the IAS. In order to verify the universal magnetic response of the S-ICD, six different magnets and one smartphone with MagSafe technology were tested. All magnet models suspended arrhythmia detection and IAS, while the smartphone did not cause magnet interferences. Conclusions: This report demonstrates the safety and efficacy of all clinical magnet models in inhibiting IAS. In case of pseudo-ES, any type of magnet allows ED providers to easily and rapidly disable the functionality of the devices when appropriate.
doi:10.2478/jce-2022-0001 fatcat:2ouyjj5g3ndljc6po7jd5brxxi