Referee report. For: Deriving alpha angle from anterior-posterior dual-energy x-ray absorptiometry scans: an automated and validated approach [version 2; peer review: 2 approved]

Pim Van Klij
2022
Alpha angle (AA) is a widely used imaging measure of hip shape that is commonly used to define cam morphology, a bulging of the lateral aspect of the femoral head. Cam morphology has shown strong associations with hip osteoarthritis (OA) making the AA a clinically relevant measure. In both clinical practice and research studies, AA tends to be measured manually which can be inconsistent and time-consuming. Objective: We aimed to (i) develop an automated method of deriving AA from
more » ... ior dual-energy x-ray absorptiometry (DXA) scans; and (ii) validate this method against manual measures of AA. Methods: 6,807 individuals with left hip DXAs were selected from UK Biobank. Outline points were manually placed around the femoral head on 1,930 images before training a Random Forest-based algorithm to place the points on a further 4,877 images. An automatic method for calculating AA was written in Python 3 utilising these outline points. An iterative approach was taken to developing and validating the method, testing the automated measures against independent batches of manually measured images in sequential experiments. Results: Over the course of six experimental stages the concordance correlation coefficient, when comparing the automatic AA to manual measures of AA, improved from 0.28 [95% confidence interval 0.13-0.43] for the initial version to 0.88 [0.84-0.92] for the final version. The
doi:10.21956/wellcomeopenres.20024.r51583 fatcat:o5rag55nfjhp7behatciks4u6a