A copy of this work was available on the public web and has been preserved in the Wayback Machine. The capture dates from 2018; you can also visit the original URL.
The file type is application/pdf
.
Primary leaflet tear without dystrophic calcifications in a 25 years old Ionescu-Shiley pericardial valve in mitral position
2015
Case Studies in Surgery
Despite its excellent hemodynamic performance, the Ionescu-Shiley bovine pericardial bioprosthesis was removed from clinical use in 1987 because of an unacceptable high incidence of primary structural deterioration of the valve, mainly related to its design. Few examples of very long-term durability of this prosthesis have been recently reported. In all these cases, the valve dysfunction was caused by primary severe dystrophic calcification and/or pannus formation. In contrast, we describe the
doi:10.5430/css.v1n1p8
fatcat:47kx3ekdmbhsnpsnepnxb72shm