Retinal vessel oximetry-calibration, compensation for vessel diameter and fundus pigmentation, and reproducibility

Martin Hammer, Walthard Vilser, Thomas Riemer, Dietrich Schweitzer
2008 Journal of Biomedical Optics  
The purpose of this study was to measure the hemoglobin oxygenation in retinal vessels and to evaluate the sensitivity and reproducibility of the measurement. Using a fundus camera equipped with a special dual wavelength transmission filter and a color chargecoupled device camera, two monochromatic fundus images at 548 and 610 nm were recorded simultaneously. The optical densities of retinal vessels for both wavelengths and their ratio, which is known to be proportional to the oxygen
more » ... e oxygen saturation, were calculated. From 50-deg images, the used semiautomatic vessel recognition and tracking algorithm recognized and measured vessels of 100 m or more in diameter. On average, arterial and venous oxygen saturations were measured at 98± 10.1% and 65± 11.7%, respectively. For measurements in the same vessel segments from the five images per subject, standard deviations of 2.52% and 3.25% oxygen saturation were found in arteries and veins, respectively. Respiration of 100% oxygen increased the mean arterial and venous oxygen saturation by 2% and 7% respectively. A simple system for noninvasive optical oximetry, consisting of a special filter in a fundus camera and software, was introduced. It is able to measure the oxygen saturation in retinal branch vessels with reproducibility and sensitivity suitable for clinical investigations.
doi:10.1117/1.2976032 pmid:19021395 fatcat:hwxvge72brfjjav6g3w4h56gle