Optical Micro-Angiography (OMAG) Based Range Optimization Method for Liquid Flow Rate release_iebhixk6fngdtfkcj4c36soy4i

by Peng Chen, Lei Jiang, Haixia Wang, Xiaorui Guo, Yilong Zhang, Yonghong He, Ronghua Liang

Released as a post by MDPI AG.

2018  

Abstract

Optical micro-angiography (OMAG) is a new method of detecting flow rate and widely used for in vivo imaging. Although OMAG can distinguish between flowing and stationary parts, it cannot obtain accurate flow rate information. This study proposed a range formula for OMAG and the ultrahigh-sensitivity OMAG (UHS-OMAG) method to quantify the measurement range of an entire system. The parameters of the angle between beam scanning and flow directions, the angular velocity of the galvanometer, and the offset of incident light were introduced, and a formula for calculating the range was derived. Experiments were conducted to measure fine and ultra-fine flow rates by using OMAG and UHS-OMAG methods. The minimum measured flow rate was approximately 30 μm/s, and the maximum measured flow rate was approximately 8 mm/s. Experimental results are in good agreement with the preset results.
In application/xml+jats format

Archived Files and Locations

application/pdf   726.2 kB
file_isywa4n36ndhzbznhpusqrukmm
www.preprints.org (publisher)
web.archive.org (webarchive)
Read Archived PDF
Preserved and Accessible
Type  post
Stage   unknown
Date   2018-11-26
Work Entity
access all versions, variants, and formats of this works (eg, pre-prints)
Catalog Record
Revision: 723e9fc4-603f-4242-8941-63fb2c986718
API URL: JSON